TIME OUT, waiting for pre-heating. The autoclave is
A600
not capable of reaching the pre-heating temperature.
A650
High temperature during sterilisation cycle.
A700
Pressure TIME OUT. After 40 minutes the chamber has
not reached the pressure which is necessary for the
carrying out of the sterilisation cycle.
A750
Low temperature during sterilisation cycle.
A800
No sterilisation chamber pressure release.
A900
General alarm:
Possible causes:
1. No power supply.
2. Voltage reduction greater than 10%.
The errors (E....) are indicated on the display for 4 seconds and are accompanied by a sound signal. They
are often caused by the a mistake being made by the user.
The alarms or faults (A....) are accompanied by sound signals. They remain on the display and are stored by
the microprocessor. Reset alarms by pressing the SELECT, DOOR and TECHNICAL buttons for 3 seconds
at the same time.
MEDILINE ITALIA S.r.l. SHALL NOT BE LIABLE FOR ANY TECHNICAL REPAIRS CARRIED OUT BY
UNAUTHORIZED PERSONNEL.
Reset the alarm by pressing the
SELECT, DOOR and TECHNICAL
buttons for 3 seconds at the same
time. Repeat sterilisation. If the
display keeps showing the same
error message, call for technical
assistance.
Reset the alarm by pressing the
SELECT, DOOR and TECHNICAL
buttons for 3 seconds at the same
time. Wait 10 minutes with the door
opened and let the autoclave cool.
Dry all instruments and repeat
sterilisation. If trouble persists, call
for technical assistance.
Reset the alarm by pressing the
SELECT, DOOR and TECHNICAL
buttons for 3 seconds at the same
time. Dry all instruments. Check and
eventually adjust the door as per
§15. Repeat sterilisation. If trouble
persists, call for technical
assistance.
Reset the alarm by pressing the
SELECT, DOOR and TECHNICAL
buttons for 3 seconds at the same
time. Dry all instruments and repeat
sterilisation. If trouble persists, call
for technical service.
Reset the alarm by pressing the
SELECT, DOOR and TECHNICAL
buttons for 3 seconds at the same
time and dry all instruments. Check
all drainage pipes (C, B, F on picture
A § 7) are not obstructed and that
drainage is
container. Repeat sterilisation. If the
display keeps showing the same
error message, call for technical
service.
Reset the alarm by pressing the
SELECT, DOOR and TECHNICAL
buttons for 3 seconds at the same
time and dry all instruments before
repeating sterilisation. If the display
keeps showing the same error
message, call for technical service.
48
not into
a sealed