Manual Filling Of Water Tank; Emptying The Tank - Tecno-gaz EUROPA B 24 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

PAR. 11
If the desalter does not work, the tank can be manually filled by the operator as follows:
1. Turn the autoclave on.
2. Disconnect the safety bleeder F (Picture A - § 7.)
3. Connect the funnelled pipe in the fittings provided with the F-joint (Photo A - § 7)
4. Pour distilled water in through the funnel (check for tank capacity § 3) until the MAX Led lights up.
QUALITY SPECIFICATIONS AS PER DIN EN STANDARD 285
Residual vapour
Silicium oxide
Iron
Cadmium
Lead
Heavy metal residual (expect iron, cadmium and lead)
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20° C)
Ph-value (degree of acidity)
Appearance
Hardness (E Ions of alkaline earth)
PAR. 12
The tank is drained at the time of maintenance operations.
To discharge the tank, proceed as follows:
Connect the supplied pipe to the discharge tap (Photo A - § 7).
Unscrew the discharge tap in counter clockwise direction.
Discharge waste water.
Re-tighten discharge tap without forcing it and check it is properly closed.
MANUAL WATER TANK FILLING
CEN STANDARD DIN EN 285

EMPTYING THE TANK

41
10
1
0.2
0.005
0.05
0.1
2
0.5
15
5 to 7
Colourless, clear and free from deposits or
sediments
0.02
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
µs/cm
Mmol/l

Publicité

Table des Matières
loading

Table des Matières