Gas bottle connection

The bottle must be connected to the coupler ( 52 ) on the back of the
analyser.
NOTA BENE:
It is compulsory to adjust the sample gas flow, by means of a
reducer/adjuster, at a value of 5 litres per minute in free air before
connecting the pipe to the connection ( 52 ), in order to prevent too a high
pressure from damaging the analyser.
11.3 -

Calibration

The
calibration
(through standard gas) that the analyser does not carry out correct
measurements.
A periodical calibration too can be necessary if it is requested by the rules
that are now in force.
Set the data relating to the values of the gases being contained inside the
standard gas bottle (see chapter 11.1) following the GAS BOTTLE DATA
CALIBRATION procedure.
The calibration can only be carried out by the AUTHORIZED
TECHNICAL SERVICE .