AUTOMATIC CONTROLS_________________________________________________________________ STARGAS
In such a condition carry out the following operations:
Make sure that the probe pipe is not obstructed.
Make sure that the external precleaner is not obstructed.
Make sure that the condensate separator is not dirty.
Replace the standard cartridge/filter.
NOTA BENE:
For a correct reading of the flow, let the hose always attached (6 m).
10.3 -

Supply Voltage

The supply voltage is very important for the analyser correct operation.
Please preset suitable systems indicating a wrong supply.
In case the voltage exceeds the allowed limits ( ± 10% of the supply voltage
value), the error is signalled by the displaying of two messages:
MAINS VOLTAGE LOW (BATTERY VOLTAGE LOW)
MAINS VOLTAGE HIGH (BATTERY VOLTAGE HIGH)
In case one of these messages is displayed while the test is being carried
out, it is necessary to repeat the test when the machinery supply voltage is
regular again.
NOTA BENE:
If the supply voltage should keep being unstable or exceeding the allowed
thresholds, it is suggested to use a mains voltage stabiliser (provided with
earthing), suitable for the specific features that are written on the approval
plate which is located on the back side of the analyser.
46
VACUUM HIGH
OPERATING INSTRUCTIONS