STARGAS _________________________________________________________________ AUTOMATIC CONTROLS
10.0 - AUTOMATIC CONTROLS
The machinery automatically checks the state of the parameters being
necessary for a good operation.
10.1 -
Pump off
The disabling of the pump is displayed on the LCD in the CONTROLS page
with the relevant symbol (see chapter 4.1).
•
Select the PUMP icon from the CONTROLS page and press ENTER to
enable the pump again (see chapter 4.1).
10.2 -
Flow Control
The machinery automatically controls the gas flow by checking the vacuum
value inside the suction circuit.
Insert the probe into the analyser so as to guarantee that the
flow control is carried out correctly .
In such a condition carry out the following operations:
•
Carry out a leakage test.
•
Check if the ring nut of the water trap is tightened.
•
Check if the ring nut of the standard cartridge/filter is tightened.
•
Verify if the O-Rings are in their housings and in good conditions.
OPERATING INSTRUCTIONS
VACUUM LOW
45