By default all interfaces are live at all times. It is however possible to configure the privileges for a
particular interface to either "read only" or "no access" from the Web page interface. The status
of the interface privileges may be viewed, but not changed, from the front panel.
To view the interface privileges the CONTROL mode must first be set to 2 (i.e. Output 2 only
selected). Press SHIFT, Bus Type to first display the status of the GPIB interface. To view the
privilege status of another interface, scroll through the interface list using the Jog control until the
required interface is displayed. The status will be shown as either
Remote/Local Operation
At power-on the instrument will be in the local state with the REM indicator off. In this state all
keyboard operations are possible. When the instrument is addressed to listen and a command is
received the remote state will be entered and REM will be turned on. In this state the keyboard is
locked out and remote commands only will be processed. The instrument may be returned to the
local state by pressing the LOCAL key; however, the effect of this action will only remain until the
instrument is addressed again or receives another character from the interface, when the remote
state will once again be entered.
RS232 Interface
RS232 Interface Connector
The 9-way D-type serial interface connector is located on the instrument rear panel. The pin
connections are as shown below:
Pin
1
2
3
4
5
6
7
8
9
RS232 Connections
The RS232 interface should be connected to a standard PC port using a fully wired 1:1 male-
female cable without any cross-over connections. Alternatively, only pins 2, 3 and 5 need be
connected to the PC, but with links made in the connector at the PC end between pins 1, 4 and 6
and between pins 7 and 8, see diagram.
Name
Description
RI
Passively asserted (+V through 10kΩ)
TXD
Transmitted data from instrument
RXD
Received data to instrument
CTS
GND
Signal ground
RTS
Passively asserted (+V through 10kΩ)
DSR
No internal connection
DTR
CD
No internal connection
Full
,
oFF
or
rEAd
.