9. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries.

10. Technical specifications

Model
BC 85
Type
BC 80
Measurement method Oscillometric, non-invasive blood pres-
sure measurement on the wrist
Measurement range
Cuff pressure 0-300 mmHg,
Systolic 50-250 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
Max. permissible standard deviation
curacy
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 97 mm x W 68 mm x H 22 mm
Weight
Approximately 108 g (without batteries)
Cuff size
Permissible operating
conditions
Permissible storage
conditions
Power supply
Battery life
Accessories
Classification
Data transfer
The serial number is located on the device or in the battery
compartment.
• This device is in line with European Standard EN 60601-
1-2 and is subject to particular precautions with regard to
electromagnetic compatibility (EMC). Please note that port-
able and mobile HF communication systems may interfere
with this unit. More details can be requested from the stated
29
135 to 230 mm
+5 °C to +40 °C, 15-93 % relative hu-
midity (non-condensing)
-25 °C to +70 °C, < 93 % relative humid-
ity, 700-1060 hPa ambient pressure
2 x 1.5V
AAA batteries
For approx. 300 measurements, de-
pending on levels of blood pressure
and pump pressure
Instructions for use, 2 x 1.5V AAA bat-
teries, storage box
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
2402MHz – 2480MHz frequency band
Transmission power max. 0 dBM
The blood pressure monitor uses Blue­
®
tooth
low energy technology
Compatible with Bluetooth 4.0 smart-
phones/tablets
List of supported smartphones/tablets