Télécharger Imprimer la page

IKA LR 1000 basic Mode D'emploi page 22

Masquer les pouces Voir aussi pour LR 1000 basic:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
Menu structure:
Stirring
Torque trend measurement ............................................................... -
Intermittent mode
Speed limit ............................................................................. 150 rpm
Heating
Control method
Limits
Temperature sensor
Weighing
Measurement .................................................................................. -
Calibration
Reset calibration values......................................................................... -
Timer
Set............................................................................................... 00:00:00 [hh:mm:ss]
Menu
Display......................................................................................... -
pH Probe
Calibration
pH-Sonde zurücksetzen .................................................................. -
Display .......................................................................................... -
Operating Mode
A.................................................................................................. activated
B................................................................................................... -
C................................................................................................... -
Display
Torque trend measurement................................................................ -
pH value....................................................................................... -
Timer ........................................................................................... activated
Axis allocation
Graph
Axis Scaling
Programs
Program 1...................................................................................
...
............................................................................................... -
Program 5...................................................................................
Modules
Valve
Safety
Time out
Password .......................................................................................... 000
Safe Temp. confirmation ........................................................................ -
Languages
Settings
Display
Sound
Factory settings
............................................................................... -
Bluetooth
..................................................................................... -
Information
42
Factory settings
Run/Stop ........................................... -
Interval
Run Time...............
01:00 [mm:ss]
Stop Time.................. 00:10 [mm:ss]
Accurate ............................................... activated
Fast ....................................................... -
Medium limit .......................................... 120
Limit information
Medium limit ................. 120
Heating block ............... 180
Calibration
2-point calibration ......... -
3-point calibration........... -
Reset calibration values ............................... -
2-point calibration..................................... -
3-point calibration .................................... -
2-point calibration ................................... -
3-point calibration ................................... -
Temperature ............................................. -
Speed......................................................... -
Torque..................................................... -
pH...................................................................... -
Automatic................................................ activated
Manuel............................................................... -
-
-
ON..................................................................... deactivated
Default Status
Open.............................. -
Closed............................. activated
Temperature (Int)
Higher.............................. -
Lower............................... -
Int > Set Temperature...... deactivated
Set ...................................................... 00:30 [mm:ss]
Speed .................................................... 10 rpm
Temperature ........................................... 30
English................................................. activated
German................................................ -
···
Background
Black ......................... activated
White ....................... -
Key tone................................................. -
Version .................................................... -
Operating mode .................................... -
Interval run ............................................. -
Interval stop .......................................... -
Medium limit ........................................... -
Heating block Limit ................................ -
Time out .................................................. -
Time out temperature .............................. -
Time out speed ....................................... -
Menu (Details):
Stirring:
Torque trend measurement:
The torque trend measurement is used to deduce the
change in viscosity of the reaction medium. The device is
not designed to measure absolute viscosity. It only measure
and display the relative change in the viscosity of the me-
dium from a starting point specified by the user.
The value can always be reset to 0% by pressing the "Back"
(T, see Fig. 12) membrane key.
Note: Torque trend measurement only works for a con-
stant set speed for the duration of the measurement.
As a result, intermittent mode cannot be used in conjunction
with torque trend measurement.
The current control variable is saved as the reference 0%,
ΔP and shown on the digital display. The change in the vis-
cosity is then shown in %. Depending on whether the vis-
cosity increases or decreases, the percentage rises or falls
above or below 0% respectively.
Intermittent mode:
The menu allows the user to activate the "Run/Stop" func-
tion. The run time and stop time can be set separately.
Speed limit:
The menu allows the user to set the desired maximum up-
per speed limit for the reactor system. The initial setting is
the maximum permissible speed of the stirrer. If the user
changes this setting, control system of the reactor saves the
new value for future stirring tasks.
If the "Speed Limit" has been changed, then the speed can
be adjusted only within the new range.
Heating:
Control mode:
In the menu, the user is allowed to select "Accurate" or
"Fast" control mode by rotating and pressing knob (A). The
selected control mode is indicated by a tick.
Fast: reach target temperature quickly but with big over-
shoot and large hysteresis at the beginning.
Accurate: Reach the tar-
T
get
temperature
takes
Set
somewhat longer, but for
this reason, the initial over-
shoot and the hysteresis
are significantly smaller.
Limits:
In "External (ext)" option,
the user is allowed to set
the maximum and mini-
mum temperature for ex-
ternal temperature control.
Confirm and store the set-
ting by pressing on "OK".
Temperature sensor:
In "Calibration" option, the user is allowed to calibrate the
external temperature sensor.
Weighing:
Measurement:
With the weighing function, the user can perform simple
weighing tasks.
Note: The heating and stirring functions must be deactivated.
Calibration:
• Open the "Calibration" submenu and confirm by pressing
the rotary knob (A, see Fig. 12)
• Enter the calibration weight and confirm by pressing the
rotary knob (A, see Fig. 12).
• Place the calibration weight on the device and wait until
the device indicates that the calibration process has been
completed.
Depending on the selected calibration mode, 2-point or 3-point
must follow these steps two or three times. Once the calibra-
tion process has been completed successfully, the weighing
module is ready for use. Regularly re-calibrate the device.
Timer:
In the menu the user can specify that the timer is displayed
on the working screen. A tick shows that the option is acti-
vated. This setting allows the user to specify the actual time
for the heating procedure. A default time can also be set
for the timer. This setting allows the user to start the heating
task for a standard time. The device stops automatically after
expiry of the set time, and the set time used for the heating
procedure appears in the display.
Note: The user can stop the stirring function before expiry
of the set time. In this case the countdown of the timer is
interrupted.
pH Probe:
Calibration:
The pH Probe must be calibrated before being used to at-
tempt a pH measurement. The calibration is used to adjust
the pH probe and the device so that they work together cor-
rectly. As part of the process, the neutral and pH gradient are
specified for the measurement chain. To complete the cali-
bration, use buffer solutions in accordance with DIN 19266.
Note: a pH calibration can only be carried out with an in-
serted temperature sensor.
Reset pH probe:
Reset the pH measuring reference.
t
Display:
Display the measured pH value on working screen.
Operating Mode:
Mode A:
After power on/power failure no automatic restart of func-
tions.
Mode B:
After power on/power failure automatic restart of func-
tions, depending on previous settings.
Mode C:
Set values (set in A or B) cannot be changed. After power
on/power failure automatic restart of functions, depending
on previous settings.
43

Publicité

loading

Ce manuel est également adapté pour:

Lr 1000 control