Part E -Maintenance And Routine Care; Section E1: Preventative Maintenance Service; Section E2: Purge Cycle; E2.1 When To Run A Purge Cycle - Nanosonics trophon 2 Mode D'emploi

Masquer les pouces Voir aussi pour trophon 2:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 63
To access the records, select:
Menu
Records then choose to view one of the record list options detailed below and follow the on
screen instructions.
Last Cycles
View most recent HLD cycles on screen and optionally print the details of a selected cycle.
Disinfectant *
Disinfectant cartridges scanned and used by the trophon2.
Chemical Indicator *
Chemical Indicator batches logged and used by the trophon2.
Probe *
Probes scanned for disinfection in the trophon2.
Operator *
trophon2 Operators logged by scanning their Operator Card.
NOTE: The Operator ID listed (if AcuTrace enabled) is the user responsible for logging the pass or fail
Chemical Indicator result via the Operator Card at the end of HLD cycle.
USB Export
Disinfection cycle records can be downloaded onto USB. Insert a USB drive into any of the three
ports on the left hand side of the trophon2, then select Download records to USB and follow the on
screen instructions.
*AcuTrace must be enabled to view this record
PART E –MAINTENANCE AND ROUTINE CARE

SECTION E1: Preventative Maintenance Service

Your trophon2 requires annual preventative maintenance and a major service at every 5,000 cycles.
Once the service due reminder screen message is displayed, please contact your customer service
representative to arrange a service. Contact details, if not provided by your distributor can be found on
the Nanosonics website.
The service options found: Menu
should only be used when advised by service personnel.

SECTION E2: Purge Cycle

The purge cycle removes and converts any remaining disinfectant from the trophon2, into oxygen and
water.

E2.1 When to Run a Purge Cycle

Prior to transporting the trophon2, if it has previously been in use.
When an on screen message states that your trophon2 requires purging. This will occur upon
disinfectant expiry at 30 days after insertion. Follow the on screen instructions to purge.
NOTE: Purging can be deferred until convenient on the day of expiry.
Initiate a purge cycle:
When the trophon2 detects an error that requires a service call.
Before lifting or moving the trophon2.
For troubleshooting purposes when directed by service support only.
L03065 2.0 9/2019
Maintenance are used to troubleshoot device malfunctioning and
19