Nanosonics trophon 2 Mode D'emploi page 18

Masquer les pouces Voir aussi pour trophon 2:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 63
5. Ensure that the probe is correctly positioned in the chamber. The probe must not make contact
with the chamber wall and must be positioned at or above the embossed line at the bottom of the
chamber.
6. If the probe is disconnected from its connector end, the external portion of the cable and
connector can be secured by placing the connector carefully inside the cable tray.
NOTE: Incorrect positioning of the probe may result in:
 Unsuccessful HLD cycles.
 Residual disinfectant remaining on the probe's surface. This may lead to temporary bleaching
and / or irritation of the skin if gloves are not worn.
 Damage to the probe may occur if in contact with the chamber wall.
NOTE: Curved probes must be correctly inserted in the trophon2. See Figure 12.
L03065 2.0 9/2019
Figure 10
Figure 12
Figure 11