Ottobock 1C66 Triton smart ankle Instructions D'utilisation page 28

Masquer les pouces Voir aussi pour 1C66 Triton smart ankle:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 34
Button functions
Main switch (recessed, black button to the left of the LED)
Function
Turn on the prosthetic foot Press briefly
Turn off the prosthetic foot Hold until red LED is off
Function
Standby mode on/off
Heel height adjustment
Erase user Bluetooth link
6.3.1 Standby mode
Standby mode can be used to save energy. In standby mode, the prosthetic foot assumes a neut­
ral position and no longer adapts to various conditions. The Bluetooth connection of the prosthet­
ic foot is turned off as well.
► Briefly press the selection key.
→ Activate standby mode: A short beep is emitted and the LED flashes yellow.
→ Deactivate standby mode: A short beep is emitted and the LED flashes green to indicate the
charge level (5 times when full).
6.3.2 Setting the Heel Height
If the prosthesis is not being worn, the heel height of the prosthetic foot can also be adjusted
without the Galileo™ app.
Prerequisite: The prosthesis is not being worn.
>
1) Put on the chosen shoe.
2) Set down the prosthetic foot with full contact and no load (see fig. 11).
3) Press and hold the selection key for about 3 seconds until the LED lights up green.
→ Measuring the position takes 5 seconds. A beep is emitted at the beginning and after
every second.
4) Do not move the prosthetic foot during the measurement.
→ A beep is emitted after the end of the measurement:
1-second beep: The measurement was successful.
3 short beeps: The measurement failed and must be repeated.
28
Button press
Selection key (white button to the right of the LED)
Button press
Press briefly
3 to 5 seconds
Min. 8 seconds
Prosthetic foot feedback
LED flashes green to indicate the
charge level (5 times if full); 1 brief
beep
If the charge level is low, the LED
lights up yellow for 1 second, 1 brief
beep
LED lights up red until the prosthetic
foot is turned off.
When the prosthetic foot is turned off,
the heel height is reset to the value that
the orthopaedic technician set during
the bench alignment.
Prosthetic foot feedback
Standby on: 1 short beep, LED briefly
flashes yellow
Standby off: 1 short beep, LED
flashes green to indicate the charge
level (5 times when full)
1 second beep + 5 short beeps (count­
down)
When the button is pressed long
enough, the LED lights up green.
1 second beep, LED briefly flashes red

Publicité