The electrodes have to be positioned to form a square over the painful zone by
using the channel 1 and 2 as shown above in Figure 1 (red or black up or down
are not important for the therapy purposes, follow the indications in the
Positions manual). The intensity should be adjusted to a level between the
thresholds of perception and pain: the maximum intensity level is the moment
in which the muscles surrounding the treated area begin to contract; over this
limit the stimulation does not become more effective, just more irritating, so it
is best to stop before that point.
Programs specifications
Medical
Prg
prg.
Yes/No
1
Yes
2
Yes
3
Yes
4
Yes
5
Yes
IACER Srl
Description
Conventional
Tens (fast)
Endorphinic
Tens (delayed)
Tens at
maximum
values
Anti-
inflammatory
Neck
pain/headache
PHASE 1
PHASE 2
Total time 40
min
frequency 90
Hz
impulse
width 50µs
Total time 30
min
frequency 1
Hz
impulse
width 200µs
Total time 3
min
frequency
150 Hz
impulse
width 200µs
Total time 30
min
frequency
120 Hz
impulse
width 40µs
Total time 20
Total time
min
5 min
92
PHASE 3
Total
time 10
min
MNPG154-06