Electromagnetic
compatibility tables
The I-TECH PHYSIO equipment has been designed and manufactured according
to the TECHNICAL STANDARD on ELECTROMAGNETIC COMPATIBILITY
legislation EN 60601-1-2:2015 with the aim of providing adequate protection
from harmful interference when installed in homes and health establishments.
The equipment does not generate significant radio frequency energy and is
adequately immune to radiated electromagnetic fields. Therefore, it does not
detrimentally interfere with radio-electric communications, electro-medical
equipment for monitoring, diagnosis, therapy and surgery, office electronic
devices such as computers, printers, photocopiers, fax machines, etc. or any
electric or electronic equipment used in these environments, as long as the
equipment complies with the ELECTROMAGNETIC COMPATIBILITY directive.
In general, the use of accessories other than those specified or provided by the
manufacturer could result in increased electromagnetic emissions or
decreased electromagnetic immunity of the I-TECH PHYSIO and result in
improper functioning.
In any case, in order to avoid any interference problems, it is recommended to
use the therapy equipment enough far away from critical equipment for
monitoring vital patient functions, and to be careful when applying therapy to
patients with pacemakers. In any case it is recommended to use the equipment
at least at 3 meters away from televisions, monitors, cellphones or any other
electronic equipment, in particular portable RF communications equipment
(including peripherals such as antenna cables and external antennas) should
not be used closer than 30cm (12 inches) to any part of the device, including
the cables specified by the manufacturer; otherwise, it could lead to
degradation of the performance of the I-TECH PHYSIO.
In conclusion, the use of I-TECH PHYSIO adjacent to or stacked with other
equipment should be avoided, since it could cause improper functioning. If
such use is necessary, the I-TECH PHYSIO and the other equipment should be
constantly observed to verify that they are operating normally.
When I-TECH PHYSIO is used in an environment relatively dry, strong
electromagnetic interferences usually occur. At this time, the device may be
affected as follows:
the device stops supplying;
the device turns off;
the device restarts.
The above phenomena do not affect the basic safety and essential
performance of the device, which can be normally used according to the