13.2.4
Conductivity measuring amplifier, electrodeless
Measuring ranges
Accuracy
Resolution
Temperature compensation
Calibration
13.2.5
Conductivity measuring amplifier, contact
Measuring ranges
Accuracy
Resolution
Temperature compensation
Calibration
13.2.6
Standard signal module
Signal input terminal 1
Signal input terminal 2
Selection of display units
Selection of measuring ranges
13.2.7
Chlorine measurement amplifier
Measuring area:
Accuracy:
Resolution:
Calibration:
13.3
Controller functions
13.3.1
Signal unit
Setpoint value (W):
Switch difference (XSD):
ON delay:
OFF delay:
Switching point distance (LW):
Switch difference (X2SD):
417101163_Multronic.doc
0-2 mS, 0-20 mS, 0-200 mS, 0-2000 mS
configurable
< 1% of measuring range end value
1 - 0.001 mS, depending on measuring range
Manual: 0 - 100 °C
Automatic with NTC: 0 - 100 °C
Reference temperature: 25 °C
Temperature unit °C or °F
with calibrating box, depending on measuring range
Auto-read function for stable measured value
0-2 µS, 0-20 µS, 0-200 µS, configurable
< 1% of measuring range end value
1 - 0,001 µS, depending on measuring
Manual: 0 - 100 °C
Automatic with PF100: 0 - 100 °C
Reference temperature 25 °C
Temperature unit °C or °F
with calibration box, depending on measuring range
Auto-read function for stable measured value
0(4)-20 mA (Option: galvanic seperation)
0(4)-20 mA (Option: galvani seperation)
pH, redox (mV), mA, mV, conductivity,
Celsius, Fahrenheit
Redox: 0 - 1000 mV
Celsius: 0 - 100 °C
Fahrenheit: 32 - 212 °F
0..2 mg/l, 0..20 mg/l
< 1 % from the measurement area value
0.1 mg/l to 0.01mg/l depending on the measurement
range
One point calibration with reference measurement
Measuring module range
0 .. 30.0 %
0 .. 240 seconds
0 .. 240 seconds
± Measuring range
0 .. 30.0 %
- 82 -
range
Rev: 5.07-08