Installation; Unit Installation; Wall Mounting; Installation In Control Cabinet - Ecolab MULTRONIC Notice Technique

Appareil de mesure et de régulation
Masquer les pouces Voir aussi pour MULTRONIC:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5

Installation

The unit can be wall-mounted or installed in the control cabinet without using any additional
components.
5.1

Unit installation

- Dry room
- Ambient temperature 0 - 45 °C
- Installation site which is free of vibrations
ATTENTION
5.1.1

Wall mounting

Installation:
1. Open housing by loosening the 4 cover screws.
2. Scribe the four holes required. (Use the housing as a template).
3. Drill the holes using 6 mm drill bit.
4. Insert dowel plugs supplied with unit.
5. Attach housing using the screws provided.
6. Close the housing by tightening the 4 cover screws
ATTENTION
5.1.2

Installation in control cabinet

Installation:
1. Create a 213 x 259 mm (wx h) opening in the control cabinet.
2. Scribe the holes as per the drawing.
3. Drill the holes using a 4.5 mm drill bit.
4. Insert the unit into the opening from the front.
5. Fasten the housing from the back using M4 x 15 screws.
ATTENTION
417101163_Multronic.doc
Failure in observing the installation requirements can result in damage of the unit.
When closing the housing, ensure that the seal is not damaged and that it sits exactly in the
seal groove.
Using screws longer than those prescribed will damage the housing.