Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Contact TensCare customer service on +44 (0) 1372 723 434. Our staff are
trained to assist you with most issues you may have experienced, without the
need to send your product in for service.
European Medical Device Regulation requires that any serious incident that
has occurred in relation to this device should be reported to the manufacturer
and the competent authority in your country. This can be found at:
https://ec.europa.eu/docsroom/documents/36683/attachments/1/translatio
ns/en/renditions/pdf

22. GENERAL SPECIFICATION

Waveform
Amplitude (> 500 Ohm load)
Max intensity
(across a load resistance of
500Ω)
Output plug
Channels
Battery
Weight
Dimensions
Safety Classification
Software version
Applied Part
Environmental Specifications:
Operating:
Storage (≤ 3 month):
Asymmetrical rectangular
99.0 mA +/- 10%
50V zero to peak across 500Ω
Constant voltage over 470-1500 Ohm
Constant current over 160-470 Ohm
Fully shielded
Single channel
Built-in lithium-ion battery
Model: YJ 523450-1000MaH
Manufacturer: YJ Power Group Limited
Ratings: 1000mAh, 3.7Wh, 3.7V
103g
120 x 60 x 20 mm
Internal power source.
R1
Vaginal and anal electrodes. Optional skin surface
electrode pads.
Temperature range: 5 to 40⁰C
Humidity: 15 to 90% RH non-condensing
Atmospheric pressure: 700hPa to 1060hPa
Temperature range: -10-+60⁰C (≤ 1 month)
Humidity: 45 to 75% RH non-condensing
39
-10-+45⁰C (≤ 3 month)
-10-+28⁰C (≤ 1 year)

Publicité

Table des Matières
loading

Table des Matières