Damaged springs in battery
compartment.
Low battery
Low batteries.
display
No sensation
Incorrect connection.
Intensity is not strong
enough.
Output will not
Open circuit cut-out
increase
operating.
above zero
Lead not connected to body
or faulty/damaged.
(Lead wires can break at the bend where they leave the machine giving no, or
intermittent, output).
Sudden
If you disconnect and re-
change in
connect a few minutes later,
sensation
the signal will feel quite a lot
stronger.
If the above review has failed to resolve your problem, or to report unexpected
operation or events, or to provide feedback call TensCare or your local supplier or
distributor (address on back cover) for advice.
Contact TensCare customer service on +44 (0) 1372 723 434. Our staff are trained to
assist you with most issues you may have experienced, without the need to send your
product in for service.
European Medical Device Regulation requires that any serious incident that has
occurred in relation to this device should be reported to the manufacturer and the
competent authority in your country. This can be found at:
https://ec.europa.eu/docsroom/documents/3683/attachments/1/translations/en/renditi
ons/pdf

18. GENERAL SPECIFICATION

Waveform
Amplitude
(over 500 Ohm load)
Output plug
Contact supplier.
Replace batteries.
Check that you have applied the 4 electrode
pads to ensure a complete circuit.
Increase strength.
Most users will feel something at a setting
below 10.
Ensure machine is attached to your body
correctly.
Try using the second lead wire.
Purchase replacement if necessary.
Always return strength to zero after
disconnecting the leads or electrode pads.
Symmetrical Bi-Phasic Rectangular
70.0 mA zero to peak, 50 steps in A & B, 60 steps in C
+/- 10%
Fully shielded: touch proof
24