4 Select Flow Delivery Mode
Flow Mode Button and Indicator
Press the Flow Mode button to select your desired
flow mode. Pressing this button repeatedly will
toggle you back and forth between continuous
flow and pulse dose modes. In Continuous Flow
Mode the oxygen is provided at a constant flow rate
between 0.5 and 3.0 LPM. In Pulse Dose Mode, ox-
ygen is supplied in a bolus at the beginning of each
inspiration, providing a selectable range of settings
1-9 with bolus sizes 16-192 mL. Pulse Dose can be
adjusted for sensitivity and breath bolus delivery.
Flow settings in both modes should be prescribed
for you by your physician.
5 Adjust Flow Setting to Prescribed
Level
Increase or Decrease Flow Setting
Button
Using the + or - buttons, adjust the setting to the
prescription supplied by your physician.
Continuous Flow Mode Display
Pulse Dose Mode Display
WARNING: IT IS VERY IMPORTANT TO
SELECT ONLY THE PRESCRIBED LEVEL OF
OXYGEN. DO NOT CHANGE THE FLOW SE-
LECTION UNLESS YOU HAVE BEEN DIRECT-
ED TO DO SO BY A LICENSED CLINICIAN.
THE PORTABLE OXYGEN CONCENTRATOR
MAY BE USED DURING SLEEP UNDER
THE RECOMMENDATION OF A QUALIFIED
CLINICIAN. YOUR DELIVERY SETTINGS
SHOULD BE PERIODICALLY REASSESSED
FOR EFFECTIVENESS OF THERAPY.
6 - ENG
PN 21182453-C4 F | User Manual
6 Begin Using the Eclipse 5
Air Inlet Filter
Your unit is equipped with an Air Inlet Filter (extra
filter provided). Check daily and clean once a week.
Connect the oxygen supply
tubing to the oxygen outlet and
connect the oxygen cannula
per the instructions provided
with the cannula, or per the
instructions from your home
care provider.
Breathe normally through the nasal cannula.
When operating on battery power, utilizing the
Pulse Dose Mode will achieve longer dura-
tion-of-use.
7 Universal Cart Operation
Align the locater holes on the bottom of the device
to the tabs on the Universal Cart. Align the large
threaded screw to the insert at the back of the
Eclipse 5. Then hand-tighten the knob on the cart
until the Universal Cart is tightly secured to the
device. Depress the push-button on the telescoping
Universal Cart handle to adjust the height of the
handle.
Contact your equipment provider for available
training options. If you do not know your provider,
contact CAIRE.
Note: For assistance, if needed, in setting up, or
maintaining the Eclipse 5, or to report unexpect-
ed operation or events, contact CAIRE or CAIRE
Representative.