Télécharger Imprimer la page

CAIRE Eclipse 5 Manuel D'utilisation page 23

Masquer les pouces Voir aussi pour Eclipse 5:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
Cleaning, Care & Routine Maintenance
Routine Maintenance
Servicing of the internal components inside the
cabinet of the Eclipse 5 must be conducted by a
CAIRE trained and qualified service technician as
needed.
Preventative Maintenance (PM) to be completed
on an annual or yearly basis. Consult your provider
to arrange PM schedule. Your home care provider
or qualified technician will perform inspections of
the alarms, internal 9-volt battery, battery, internal
filters and other internal parts only as needed.
User Care and Cleaning of the Device
Cannula Replacement
Replace your supply tubing and cannula on a
regular basis as recommended by your home care
provider. Your physician or home care provider
will provide you with cleaning, disinfection and
replacement information.
Note: Always follow the cannula manufacturer's
instructions for proper use. Replace the disposable
cannula as recommended by the cannula manufac-
turer or your Equipment Provider. Additional sup-
plies are available from your Equipment Provider.
Air Inlet Filter
Ambient air is drawn into the device through the air
inlet located on the top rear of the device. Cleaning
the air inlet filter is the most important maintenance
activity you will perform as a user. Cleaning the
air inlet filter will keep your Eclipse 5 performing
properly. Check this air inlet filter daily and clean it
at least once a week using the following procedure:
1. Remove the air inlet filter
2. Wash the filter in
warm tap water
using a mild
soap detergent
solution
3. Rinse the filter
thoroughly with
clear tap water
and squeeze out
the excess water
4. Allow the filter to
air dry
5. Reinsert the filter
in the cabinet
Cleaning tip: Keep a clean second filter as a re-
placement for use when one filter is drying.
CAUTION: Do not operate the Oxygen
Concentrator without the air intake
filter in place. If a second filter is pro-
vided, insert the "replacement" filter
before you clean the dirty filter. Clean
the dirty filter in a warm soap and
water solution then dry thoroughly
prior to use.
Note: The Manufacturer does not recommend the
sterilization of this equipment.
Note: Do not operate the Oxygen Concentrator
without the air intake filter in place. If a second
filter is provided, insert the "replacement" filter
before you clean the dirty filter. Clean the dirty
filter in a warm soap and water solution then dry
thoroughly prior to use.
The air inlet filter should be replaced annually. If
the Eclipse 5 is used in a dusty environment, the
filter may need to be replaced more frequently. You
should not operate the Eclipse 5 without an air inlet
filter installed for more than 5 minutes.
AC Power Supply, DC Power Supply, Power Cord
and Outer Case:
Turn OFF the Eclipse 5 and disconnect from AC or
DC power before any cleaning or disinfection ac-
tivity. DO NOT spray the outer case directly. Use a
damp (not wet) cloth or sponge. Spray the cloth or
sponge with a mild detergent solution to clean the
cabinet and power supplies. To disinfect the Eclipse
5, use Lysol® Brand II disinfectant (or equivalent).
Proceed as directed by the cleaner manufacturer.
PN 21182453-C4 F | User Manual
Eclipse 5
23 - ENG

Publicité

loading