Recommended Operating Environments
For proper use of the device, the following chart
provides important information concerning the
recommended operating environments, or operating
conditions.
Specifications
Operating
50º F to 104º F (10º C to 40º C)
Temperature
Operating
10% - 95% at an 82.4° F (28° C)
Humidity
dew point
Transport/Stor-
-4º F to 140º F (-20º C to 60º C)
age
Humidity: Up to 95% Non-Con-
Temperature
densing
Electrical
Use no extension cords. Use no
electrical outlets controlled by a
switch.
Altitude
0 – 13,123 feet (0 - 4,000
meters)
Placement
DO NOT block the air inlet (top
back cover) or the exhaust vent
(back bottom left). Place device
a minimum of 3 inches (7.5
cm) away from walls, draperies,
furniture, etc.
Environment
Must be smoke, pollutant, and
fume free.
Operating Time 24-hours a day when connected
to an external AC or DC power
source.
Oxygen
90%+5.5%/-3% @ sea level
Concentration
Dimensions
19.3 high x 12.3 wide x 7.1 deep
(inches), 49.0 height x 31.2 wide
x 18.0 deep (CM)
Weight
Eclipse 15 lbs., Battery 3.4 lbs.
Power
AC Power (100-240 VAC, 50-60
Hz; DC Power (12V nominal);
Battery (Quantity (2) 95 W-Hr
battery packs each containing
7.92 gm of lithium content)
Nominal
1.8 to 5.0 hrs recharge time to
Battery
achieve 80% capacity (dependent
Recharge Time
upon the flow)
Alarms/Alerts
Loss of power, Low Battery, Low
Therapeutic O2 Output, O2 flow
outside normal limits. No Inspi-
ration detected in Pulse Dose
Mode, Unit Malfunction
O2
Green Light=Normal
Concentration
Yellow=Caution <85%
Indicator
Outlet
Nominal: 7.0 psig
Pressure
Maximum: 14.0 psig
Sound Level
48 dBA at 3.0 LPM Continuous
Flow Mode; 40 dBA at 3.0 Pulse
Dose Setting; 59 dBA while
alarming
Applied Part
Cannula
The expected service life is a minimum of five years.
WARNING: USE OF THIS DEVICE AT AN
ALTITUDE ABOVE 13,123 FEET (4,000
METERS) OR OUTSIDE TEMPERATURE
OF 50º F TO 104º F (10º C TO 40º C) OR
A RELATIVE HUMIDITY ABOVE 95% IS
EXPECTED TO ADVERSELY AFFECT THE
FLOWRATE AND THE PERCENTAGE OF
OXYGEN AND CONSEQUENTLY THE
QUALITY OF THE THERAPY.
Note: If the Oxygen Concentrator has been stored
for and extended period of time outside its normal
operating temperature range, the unit should be
allowed to return to normal operating temperature
before use. The recommended time is 3 hours.
Available Settings
The Eclipse 5, including the cannula and other
accessories, are specified for use at the following
specific flows.
Continuous Flow Settings
in LPM
0.5
1.0
1.5
2.0
2.5
3.0
-
-
-
-
-
-
-
-
-
Minimum oxygen concentration (82%).
PN 21182453-C4 F | User Manual
Eclipse 5
Pulse Dose Setting Bolus
Size (mL)
-
16
24
32
40
48
56
64
72
80
88
96
128
160
192