Powering on your Eclipse 5
The device is capable of being
operated directly from three (3)
different power sources:
• AC Power Supply with
NEMA Power Cord
• DC Power Supply
• Rechargeable Power Car-
tridge (Battery)
General
Charging
Information
The Eclipse
5 charges the
battery while
using AC power and when the battery temperature
is within safe charging temperature. When external
power is disconnected, the device will automat-
ically switch over to the battery if the battery is
installed and charged. When external power is
restored by plugging into an AC outlet, the device
will accept power from the external power source
and replenish the battery. It is not mandatory
for the battery to be installed for the Eclipse 5 to
operate.
Proper connection of the AC or DC Power Supply
requires aligning the power plug with the recessed
external power receptacle of the device. The
recessed external power receptacle is located on
the right side of the device when facing the Control
Panel. Push the power plug into the recessed
receptacle.
WARNING: THE ECLIPSE BATTERY
CARTRIDGE AND DC POWER SUPPLY
MUST BE PERIODICALLY CHECKED.
WHEN AC POWER IS NOT AVAILABLE
THESE ALTERNATE POWER SOURCES
MAY BE USED.
AC POWER SUPPLY WITH TWO PINS
The Eclipse 5 includes a universal AC Power
Supply for use at home or wherever standard AC
power is available. To connect to an AC power
source, be sure the AC Power Supply cord is
securely inserted into the recessed power receptacle
on the side of the device and that the cord from the
AC Power Supply is plugged into an AC outlet.
When the Eclipse 5 is plugged in properly, a green
indicator on the AC Power Supply will light and the
External Power Indicator,
user control panel.
Step 1
Note: Use only electrical voltage specified on the
specification label affixed to the device.
CAUTION: Use only accessories rec-
ommended by the manufacturer. Use
of any other may be hazardous, cause
serious damage to your oxygen con-
centrator and will void the warranty.
CAUTION: Do not use extension cords with
this unit or connect too many plugs into the
same electrical outlet. The use of extension
cords could adversely affect the performance
of the device. Too many plugs into one outlet
can result in an overload to the electrical panel
causing the breaker/fuse to activate or fire if
the breaker or fuse fails to operate.
CAUTION: Use of cables and adapters other
than those specified, with the exception of
cables and adapters sold by the manufacturer
of the medical electrical equipment as replace-
ment parts for internal components, may result
in increased emissions of decreased immunity
of the Oxygen Concentrator.
PN 21182453-C4 F | User Manual
Eclipse 5
AC Two-Pin Power Supply