Alarms Table
Alarm
All OK, warm-up done
Warm-up O2 < 85%
No breath detected (less than
45 seconds)
No breath detected (exceeding
45 seconds)
One wire communication loss
Cannot charge battery
Ambient pressure fail
O2 < 85% (After warm-up)
Flow rate error > +/-10%
Warm battery > 60°C
Low battery
(5% remaining battery
capacity)
Loss of power
Malfunction
20 - ENG
PN 21182453-C4 F | User Manual
LCD Text
Green
NO ALARM
ON
O2 LOW
ON
NO ALARM
ON
P <--> C
ON
ONE WIRE
ON OR
OFF
CHARGER
ON OR
OFF
AMB PRESS
ON OR
OFF
O2 LOW
OFF
FLOWRATE
OFF
WARM BAT
ON
LOW BAT
ON
********
OFF
FAIL XX
OFF
Yellow
Audio
OFF
OFF
ON
No Beeps
OFF
OFF
ON
1 alarm beep, repeat
every 20 seconds
ON
1 alarm beep, repeat
every 20 seconds
ON
1 alarm beep, repeat
every 20 seconds
ON
1 alarm beep, repeat
every 20 seconds
ON
1 alarm beep, repeat
every 20 seconds
ON
1 alarm beep, repeat
every 20 seconds
ON
1 alarm beep, repeat
every 20 seconds
ON
1 alarm beep, repeat
every 20 seconds
ON
1 alarm beep, repeat
every 20 seconds for
200 seconds
ON
1 alarm beep, repeat
every 20 seconds for
200 seconds
Code
000
008
000
200
100
001
004
008
020
002
010
040
080