Cleaning of housing and base
Do not use pointed implements or aggressive detergents. Clean the housing and the device
base with a moist, soft cloth and a mild soapy solution. Alternatively, use a alcohol/water mixture.
To prevent residue, we recommend wiping the device dry with a soft cloth.
Before cleaning the device, disconnect the power plug from the socket!
Cleaning of TOPITEC
Step 1: After use, wipe the shaft with a dry cellulose
cloth/tissue paper (dry cleaning).
Step 2: Use a cellulose cloth which has already been soaked with a surface disinfectant (e.g. WEPA
surface disinfectant 1,000 ml with spray head, order no. 040141) or 2 propanol 70% for
thorough cleaning. You can of course also use disinfecting wet wipes such as Meliseptol
Dispose of the used wipes/tissue paper.
Step 3: Store the tool shaft in the
storage box supplied.
The tool shaft can also be cleaned in a rinsing bath, provided that
only its bottom part (mixing disc holder) is immersed in water!
After cleaning, dry the shaft.
You can find further information at:
topitec.de/ Produkte/ Werkzeugwelle/ Reinigungsempfehlung
topitec.de/ Support/ Support-Video
Statutory requirements re. cleaning
Please draw up internal cleaning and hygiene regulations that take into account the relevant
statutory requirements for pharmacies, e.g. ApBetrO, BAK Guidelines, GD Hygiene Guidelines, PSA,
trade regulations, etc.