wissner-bosserhoff sentida 7-i Notice D'utilisation page 84

Le lit de soins digital
Masquer les pouces Voir aussi pour sentida 7-i:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 118
Danger!
Liquids on the CareBoard`s
the low nursing bed. After use, always press the emergency stop button. Immedia-
tely wipe liquids off the display and also immediately dry it after cleaning.
Caution!
When mounting the CareBoard
with the bed when it is moved or adjusted.
If the CareBoard
If 10 minutes pass without any action via the CareBoard
logged out and the equipment switches to standby mode. This can also be perfor-
med manually by pressing the emergency stop button.
There is a bank of buttons to the left of the display. At the top, the bank of buttons shows the
GO safety button which activates the CareBoard
There are 3 further buttons below it, which offer access to important acute functions. This
direct access is also called acute care pack.
Difference between display off and log out: When the CareBoard
sing the GO safety button once switches off the display. The user is only logged out
if the emergency stop button is pressed.
You can attach a bracket for the comfort resident handset to the SafeFree
left or right at the head end.
This acute care pack contains the following functions:
– Automatic electrical CPR
CPR (cardiopulmonary resuscitation), this corresponds to the comfort bed egress
position with horizontal position of knee and backrest. The CPR automatic is used
during reanimation by cardiac massage.
– Automatic electric Trendelenburg
Here, you can set the Trendelenburg position. The patient surface is even and the
patient´s head is placed below the patient´s center of gravity on a surface which is
inclined at an angle of at least 12°.
– Emergency stop button
By pressing the emergency stop button, all electric motor adjustments are stopped
immediately and the CareBoard
The symbols below the bank of buttons indicate if the low nursing bed is connected to the
power supply system and if the battery is loading. The following table shows the meanings of
the battery symbol.
84
display can lead to injuries caused by movements of
®
to the wall, please make sure that it cannot collide
®
is separated from the bed, the handset also ceases to function.
®
is switched off.
®
, the user is automatically
®
; that is why this button is called GO button.
®
is activated, pres-
®
side guard on the
®