Intermittent output:
check that electrodes and cables have been connected correctly. If
the problem persists, replace cable or contact the manufacturer.
Stimulation is ineffective:
check that electrodes have been positioned correctly. If the
problem persists, contact the doctor/therapist.
Treatment zone becomes red and/or you feel a persistent pain:
the problem may be caused by the repeated treatment on the
same area: re-position the electrodes and if the problem persists
interrupt the treatment and contact the doctor/therapist.
Check that electrodes are stuck to the skin well and check that they
are not dirty and/or consumed. Change the electrodes and
continue the treatment. If the problem persists, contact the
doctor/therapist.

Battery charging

MIO-CARE is supplied by internal rechargeable Ni-MH 800mAh battery with new
long-lasting technology.
When during the treatment many intensity increases is needed or the device
turns off, it indicates a low battery state. In this case, the display shows
the middle lower side: the icon is showed only when the battery is discharged
or when the charger is disconnected from the device plug (in this case the
display shows all icons for a few seconds). Please pay attention that the
DISPLAY DOESN'T SHOW ANY BATTERY ICON FULLY CHARGED.
To proceed with the charging follow the steps below:
make sure that the device is switched off before charging with the ON/OFF
button;
make sure that the device is not being used by patient (disconnect cables
and electrodes);
connect the battery charger to the plug on the upper side of MIO-CARE and
connect the battery charger into the power socket.
After 6/8 hours the recharging will end automatically, so therefore it will be
possible to disconnect the charger from the power socket and from the device.
WARNING: at the end of the charge wait at least 30 minutes before
switching on the device; in order to allow the cooling of the battery
pack, overheated during charging and the closure of the integrated
safety system that prevents the device from turning on.
Do not immerge the battery in water or other liquids and do not expose it to heat
sources.