11 – MAINTENANCE AND PREVENTION
Correct maintenance of the autoclave assures correct functioning of it and a secure saving in terms of time
and costs for assistance and maintenance.
11.1 – BACTERIOLOGICAL FILTER SUBSTITUTION
Substitute the bacteriological filter (Fig.9 pos.2), maximum every 200 cycles, or as soon as the filter becomes
a dark color.
Use only and exclusively original TECNO GAZ filters.
FIGURE 9
Frontal view 1.2
Pos. 2
11.2 – CHAMBER CLEANING
Clean periodically the chamber, remove eventual deposits or debris, thus avoiding the introduction in the
discharging circuit of material which can cause obstructions.
For correct cleaning use only water and the abrasive sponge provided.
Never use solvents, detergents, chemical solutions, descaling agents or other similar products.
11.3 – DOOR SEAL CLEANING
Periodically remove eventual residuals which are deposited on the circumference of the seal using water and
the sponge provided (non abrasive side), or else a moistened cloth.
11.4 – DOOR REGULATION
To regulate the door lock pressure, it is necessary to adjust the regulator of the same door using the double
function wrench (extraction tray + wrench) provided.
Turn anti clockwise (fig.A) to increase lock pressure.
Turn clockwise (fig.B) to decrease the lock pressure.
FIG. A
FIG. B