8– USAGE INSTRUCTION
After having installed the autoclave, proceed with preparations and use.
8.1 – TURN ON THE AUTOCLAVE, press the general switch (fig.3 pos.16)
FIGURE 3
Frontal View 1.1
Pos. 16
8.2 – LOADING PURE WATER TANK ( lt. 4.5)
Connect the transparent hose complete with the filter, (provided), to the frontal connection of the autoclave
(fig.4 pos.3); unscrew the metal ring anticlockwise ( to execute 2 complete turns )
Insert the other end of the hose with filter to the inside of the demineralized or distilled water container.
At this point press the button
to start the water loading pump and press it for at least one second.
The pump loads the tank inside the autoclave. If maximum level is not reached within 180 seconds the pump
automatically stops, therefore it is necessary to press the button again.
The pump stops automatically when maximum level is reached.
With connection to a deionizer, if the maximum water level has not been reached, the functioning of
the autoclave will be obstructed.
FIGURE 4
Frontal view 1.2
Pos. 3
52
DXBA408 Rev_14