Patient Cable

ALIGN THE LEADS IN SUCH A WAY AS TO PREVENT ANYONE STUMBLING OVER THEM OR ANY DAMAGE
CAUSED BY THE WHEELS OF INSTRUMENT TROLLEYS. THE SAME CAUTION APPLIES TO THE STORAGE
OF THE LEADS.
The patient cable should not be exposed to excessive mechanical stress. Whenever
disconnecting the leads, hold the plugs and not the cables.
The cable can be wiped with soapy water. Sterilization, if required, should be done with gas
only and not with steam. To disinfect, wipe the cable with hospital standard disinfectant.
Cleaning the Thermal Printhead of the Internal Printer
If the printer is used a lot, a residue of printers ink ( from the grid on the paper) can build up on
the printhead. This can cause the print quality to deteriorate. We recommend therefore that
every month the printhead is cleaned with alcohol as follows:
Remove the paper tray. The thermal printhead is found under the paper tray release catch.
With a tissue dampened in alcohol, gently rub the printhead to remove the ink residue. If the
printhead is badly soiled, the colour of the paper grid ink (i.e. red or green) will show on the
tissue.
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
AT-1 User Guide Art. No. 2.510171 g
49