Confirm the selection by pressing
Note: The set value is the lower limit of the frequency range and is normally set to 0.05 Hz.
The settings 0.15 and 0.30 Hz should only be used when absolutely necessary, as
the possibility exists that they could affect the original ECG signal, especially the
ST segments.

Mains Filter

The Mains filter
Mains filter is an adaptive digital interference filter designed to suppress AC interference
Mains filter
Mains filter
Mains filter
without attenuating or distorting the ECG.
Set the mains filter in accordance with the frequency of your local mains supply as follows:
20
Mains Filter
Entry Key
Filter Setting
Sequence
5
Mains Filter 50 Hz
ALT
8
6
Mains Filter 60 Hz
9
Mains Filter Off
Confirm
Press STOP
key