Automatic Mode

Automatic Mode
Automatic Mode
Automatic Mode
Automatic Mode
a.

without external printer connected or printer switched off

In automatic mode
automatic mode
automatic mode
automatic mode, a full 12-lead ECG is printed in one of two predefined formats with selectable
automatic mode
sensitivity. These formats (see Settings) are selected by the user to suit his specific needs and
requirements.
When the AUTO GAIN key
very large waveform amplitudes and sets the sensitivity for the extremity and/or precordial leads
accordingly to reduce the overlapping of traces.
To start the automatic ECG recording in Format 1, press the AUTO START key:
To start the automatic recording in Format 2, press the ALT key followed by the AUTO START
key:
Depending upon which option is included, the printout gives the following:
• ECG recording of all leads in either Standard or Cabrera format according to selection
• Sensitivity
• Heart Rate
6
is pressed before recording in automatic mode, the unit detects
+