Switching On and Off

Switching On and Off
Switching On and Off
Switching On and Off
Switching On and Off
The AT-1 is switched on with the green ON key
and off by means of the red OFF key
The unit is automatically switched off after 5 minutes (30 seconds if battery capacity is limited)
if no key is pressed and the patient cable is not connected.

Potential Equalisation

If the AT-1 is used in conjunction with other patient connected equipment, we recommend that
the potential equalisation stud on the rear of the unit is connected to the building common
ground with the yellow/green ground cable (Part-no. 2.310005). When working from an
emergency vehicle, the vehicle common ground can be used.
THE UNIT MUST BE GROUNDED WHEN THE EXTERNAL PRINTER IS CONNECTED.
WARNING !
WARNING !
WARNING !
WARNING !
WARNING !
AT-1 User Guide Art. No. 2.510171 g
11