Patient Cable Connections

The accessory kit of the electrocardiograph includes a 10-lead patient cable. This cable is
plugged into the patient cable socket on the right-hand side of the unit and secured with the
two screws.
The CARDIOVIT AT-2 is CF rated. The patient connection is fully isolated and defibrillation
protected. Protection against defibrillation voltage is however only ensured, if the original
SCHILLER patient cable (Part-no. 2.400070 / USA: 2.400071) is used. Make sure that during
ECG recording neither the patient nor the conducting parts of the patient connection or the
electrodes (including the neutral electrode) come into contact with other persons or
conducting objects (even if these are earthed).
Page 28
CARDIOVIT AT-2
8.2001