General Settings
There are three different filters which can be set individually as follows:
•
Baseline filter
•
Mains filter
•
Myogram filter
The setting for each filter is given on the setup printout.
Baseline Filter
The digital Baseline filter suppresses excessive baseline drifts. The setting options are as
follows:
The set value is the lower limit of the frequency range and is normally set to 0.05 Hz. The
settings 0.15 and 0.30 Hz should only be used when absolutely necessary, as the possibility
exists that they could affect the original ECG signal, especially the ST segments.
Confirm the selection by pressing STOP.
Page 16
Baseline Filter
Entry Key Sequence
0
0.05 Hz (Default)
ALT
5
1
0.15 Hz
3
0.30 Hz
Filter Setting
Confirm
Press
STOP
Key
CARDIOVIT AT-2
8.2001