Directions For Use - Integra MAYFIELD 2 A3100 Mode D'emploi

Unité de base
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
EN – ENGLISH

Directions for Use

Attachment of the Base Unit to Operating Room Table:
MAYFIELD 2 Base Unit (3100) Left Bracket may be adjusted along the Connecting Tube to allow the Support Rod to
accommodate table receptacles spaced 4-1/2" to 8-1/4" (114mm to 209.5mm) apart.
MAYFIELD 2 Base Unit (3101) Left Bracket may be adjusted along the Connecting Tube to allow the Support Rod to
accommodate table receptacles spaced 5-1/8" to 8-1/4" (130mm to 209.5mm) apart.
MAYFIELD 2 Base Unit (3102) Left Bracket may be adjusted along the Connecting Tube to allow the Support Rod to
accommodate table receptacles spaced 6-5/8" to 15-3/4" (168mm to 400mm) apart.
1. Locate and loosen the Left Bracket Lock Knob on the backside (See Figure 4) of the Left Bracket and slide the Left Bracket
smoothly along the the Connecting Tube until the desired width between the Support Rods is achieved.
2. Mount the Base Unit to the operating room table by aligning the two Support Rods with receptacles at the head end of
the Operating Room table, or, in the NeuroGen Adaptor or Crossbar Adaptor.
3. Tighten the Left Bracket Lock Knob to secure the Left Bracket into position.
Do NOT use any tools to tighten any knobs on this device. All knobs should be hand tightened only. Use caution not to over-
tighten. Over-tightening may damage the device.
Clockwise rotation to tighten
4. Secure the Base Unit Support Rods to the table by tightening the table lock knobs or the knobs on the NeuroGen
Adaptor or Crossbar Adaptor.
7
CAUTION!
Figure 4 Left Bracket Lock Knob
Left bracket lock knob

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mayfield 2 a3101Mayfield 2 a3102

Table des Matières