EN – ENGLISH
MAYFIELD 2 Standard Swivel Adaptor - A3018
Disassembly
1– Remove swivel adaptor (1) from the base
unit (2).
2– Do not attempt to disassamble further.
(2) locking screw
(1) swivel adaptor
TriStar Swivel Adaptor - A3008
Disassembly
1– Completely remove swivel adaptor from
any components.
2– Unlock the TriStar bar (1), remove
Locking screw (2) and TriStar bar (1).
(2) locking screw
15
Inspecting for cleanliness
Pay close attention to:
1– Teeth of the starbursts (3).
2– Area around the locking screw (2).
3– The base of the swivel adaptor for
organic debris.
Inspecting for cleanliness
Pay close attention to:
1– Teeth of the starbursts (3).
2– Area around the locking screw (2).
3– The base of the swivel adaptor for
organic debris.
(1) TriStar bar
(3) teeth
(4) locking lever
(3) teeth
Reassembly
N/A
Reassembly
1– Re-install the dual TriStar bar (1).
2– Engage the locking lever (4) on
the TriStar bar (1).
3– Insert and firmly press the torque
screw (2) in place.