Manual Cleaning - Integra MAYFIELD 2 A3100 Mode D'emploi

Unité de base
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
EN – ENGLISH
Step 2: Cleaning
The instrument can withstand detergents having pH values between 3 and 11. Manufacturer's instructions of the selected
detergent must be followed for this device.
NOTE: The use of highly alkaline agents can tarnish and / or corrode some components. While this does not affect the
performance of the device, we recommend using an acid neutralizing agent immediately after each use to keep a device
looking new.
2 Cleanings are available:

• Manual Cleaning

• Automatic cleaning
Manual Cleaning
Following these steps is recommended:
1. Remove Base Unit from operating table.
2. Remove Swivel Adaptor from Base Unit.
3. The Base Unit should be thoroughly cleaned after each use. Scrub each component with a soft brush. The Base Unit
can withstand detergents that fall within pH ranges of 3-11. Clean thoroughly to remove any traces of blood and/or
debris and to prevent such blood or debris from interfering with function or movement. Rinse thoroughly with clean
water to remove all traces of detergent.
4. Dry all components with a soft dry towel.
5. After components are totally dry, inspect the unit for cleanliness.
6. Inspect all components to ensure that there is no visible organic debris or residue from cleaning agent. Repeat
process if any visible debris or residue is detected.