4. Instructions For Use Con't.
The buzzing noise during the air drying stage is the compressor operating. The air drying
phase of the cycle may be interrupted at any time by pressing the STOP button. The display
will read:
When the automatic 60-minute air drying stage is finished the display reads:
If a sterilization cycle is successful the reminder tone sounds and the amber light flashes
until the STOP button is pressed or the cassette is removed from the unit.

4.8 Stopping a Cycle

To stop a cycle press the STOP button. If the STOP button is pushed, the cassette is
removed, or the unit detects a problem while operating, the cycle will stop and the amber
active light will flash. Once a cycle has been stopped, the STOP button must be pressed
before another cycle can be started. The display reads any of the following messages:
If the display shows the message, CYCLE FAULT or NOT STERILE, the cassette contents
are not sterile! See section 7. Troubleshooting for more information.
If the air drying stage of the cycle is interrupted, do not store wrapped instruments that were
in the cassette unless they are dry.
PLEASE WAIT
CYCLE COMPLETE
REMOVE CASSETTE
CYCLE COMPLETE
CYCLE FAULT xxx
NOT STERILE
or
CASSETTE REMOVED
NOT STERILE
20