6. The Communication Port con't.

6.5 Cycle Printout Overview

1.
Model: STATIM 2000
software: S2S2R410
2.
Unit Identifier: autoclave has
been set up as number 323
3.
Cycle counter: the number of
cycles having been run on the
unit = 9
4.
Time / Date: 1:38 pm 14th
April, 2003
5.
Cycle Name: SOLID
UNWRAPPED (N)
6.
Cycle Name cont'd –
parameters: 134˚C /
3.5 min – 1P
7.
Cycle clock: starting at 0:00
8.
Warm up complete: start of
the conditioning phase is 2:07
(see cycle graph – 'A' phase
complete, start of 'B' phase)
9.
Peek Temp. / Press. & Time
of conditioning purge: for each
purge (number of lines based
on cycle selection – this is
a one purge cycle – activity
within 'B' phase )
10. Lowest Temp. / Press. & Time
of conditioning purge: for each
purge (number of lines based
on cycle selection – this is
a one purge cycle – activity
within 'B' phase )
1.
2.
3.
4.
5.
6.
7.
8.
9.
10 .
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
Acceptable Tolerances
Sterilization time: "Sterilization time" (e.g. 3.5 mins) -0/+1%
Saturated Steam Pressure: 304kPa - 341kPa for Unwrapped/
Wrapped cycle (205kPa - 232kPa for Rubber and Plastics cycle)
Sterilization Temperature: "Specified temp" -0/+4 (134˚C-138˚C)
(121˚C -125˚C for Rubber and Plastics cycle)
*data on Cycle Printout should fall within these ranges