EN - INSTRUCTIONS FOR USE
1 Intended use
The FlurAbsorb gas filter is intended to remove
volatile anaesthetic in operating rooms, intensive
care units, including mobile and ambulatory
treatment areas, independent of any kind of
ventilator used. The FlurAbsorb should be used
in combination with the standard holder or
mount provided to enable proper operation of
the device.
2 What is FlurAbsorb?
FlurAbsorb is a filter cartridge for removing
halogenated hydrocarbons anaesthetic gases
such as Isoflurane and Sevoflurane. The filter
consists of a labelled container with various filter
inserts, special activated carbon, a locked
closure, vents to enable air flow and a connector
which enables connection to the exhaust port of
a ventilator.
2.1 Symbols
The following symbols are used in the Instruction
for Use or on the product itself.
Symbol
Description
Indicates informa�on important for
NOTE!
op�mal use of product
Read the instruc�ons for use carefully
before use
Do not use if damaged
REF
Catalogue Number
This way up
European Medical Device Regula�on
2017/745
3. Transportation and Storage of the
FlurAbsorb gas filter
Please follow the procedures below in order to
maximise the performance of the FlurAbsorb
Filter.
3
FlurAbsorb/FlurAbsorb-S
3.1 Transportation
- Ensure the dispatch carton/box is transported
with the label on the top.
- The carton should not be put on its side or on
its top.
- Open the carton from the top.
- Do not use the FlurAbsorb filter if the seal is
damaged or broken.
NOTE!
The arrows on carton/ box is indicating
the correct position of the Carton /box.
3.2 Storage
- The dispatch carton/box should be stored with
the label on the top.
- Store in a dry dust free environment.
- Do not store in the open air.
- Store between -5°C to +35°C.
4. Operation of the FlurAbsorb Gas Filter
Before using the filter, ensure that all
components needed for use are available. These
include:
- The FlurAbsorb Gas Filter.
- The standard mount.
- Accessories.
SEDANA MEDICAL
o ers an accessory kit with
REF 26072 which includes:
- 22 mm flexible tube to connect the anaesthetic
gas filter to the exhaust of the ventilator.
- 6mm O� connecting tube/silicone tube to
connect a side-stream gas monitor with the
anaesthetic gas filter.
- Adapter 22m/22f to be used for the 6mm O2
tube for simultaneous waste of gas from the
gas monitor and the ventilator.
- Adapter 32m/30f to fit the flexible tube on
certain ventilator exhausts e.g. Hamilton
Galileo, Maquet Servo-i, Siemens Servo.
4.1 Fitting the FlurAbsorb Gas Filter on a
ventilator.
1. Attach the mount for the anaesthetic gas filter
to a standard bar on the ventilator or close to
the ventilator.
www.sedanamedical.com
2. Place the anaesthetic gas filter in the mount
and remove the protective seal cap.
3. Put the 22mm flexible tube on the exhaust of
the ventilator. For some ventilators it may be
necessary to use a 22m/30f adapter.
4. Place the 22mm flexible tube on the
anaesthetic gas filter. If a side stream gas
monitor is being used, then use a 22m/22f
adaptor with a secondary outlet to connect the
6 mm O2 tube for the exhaust from the gas
monitor.
4.2 Application with the AnaConDa /
Sedaconda ACD.
5. Considerations during filter use
- FlurAbsorb (REF 26096) Filter should be
changed when the storage capacity has
reached 400g or after 10 syringes (AnaConDa
/ Sedaconda 50ml syringes) of Anaesthetic
agent have been used.
- FlurAbsorb-S (REF 26094) Filter should be
changed when the storage capacity has
reached 125g or after 3 syringes (AnaConDa /
Sedaconda 50ml syringes) of Anaesthetic
agent have been used.
NOTE!
A space is available on top of the
FlurAbsorb (REF26096) lid to record both the
start date of use and each syringe emptied.
6. Disposal of the FlurAbsorb Filter
- Put the protective seal cap back on the filter.
- Dispose of used filter according to hospital
protocols.
EN - INSTRUCTIONS FOR USE
7. Technical Data
Standard
FlurAbsorb-S
(26096)
(26094)
Product
Class 1
Class 1
Class
Temperature Operation: +5
Operation: +5
to +40˚C
to +40˚C
Storage: -5
Storage: -5
to 35˚C
to 35˚C
Flow
<16 Pa
<34 Pa
Resistence
@10L/min
Height
22cm
13cm
Diameter
12.3cm
10.1cm
Weight
Approx. 1000g
Approx. 400g
Volume
2l
750ml
Storage
Approx. 400g
Approx. 125g
Cappacity
Product Description
REF Number
FlurAbsorb
REF 26096
FlurAbsorb-S
REF 26094
FlurAbsorb Accessory Kit
REF 26072
FlurAbsorb Mount
REF 26098
FlurAbsorb-S Mount
REF 26099
8. Warnings
- Caution when using active
humidification with FlurAbsorb.
- Do not use FlurAbsorb-S with active
humidification.
- Replace the filter immediately in case of
occlusion
Sedana Medical Ltd Address
Unit 2A, the Village Centre, Tow Mile House, Naas,
Co. Kildare, Ireland
Phone
+353 (45) 87 90 81
Email
info@sedanamedical.com
Website www.sedanamedical.com
3000 030-2107 Rev.3