Recommended safety distance
portable and mobile RF communications equipment and the device
The device is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the device as recommended below,
according to the maximum output power of the communications equipment.
Rated power of
transmitter(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distanced in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Disposal: The product must not be disposed of along with other domestic waste.
The users must dispose of this equipment by bringing it to a specific recycling point for
electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies
Recommended separation distances between
Separation distance according to power of transmitter (m)
150 k Hz – 80 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
80 MHz – 800 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
68
800 MHz – 2.5 GHz
d=2.3√P
0.23
0.73
2.3
7.3
23