Télécharger Imprimer la page

Gima ECG90A Manuel D'utilisation page 20

Masquer les pouces Voir aussi pour ECG90A:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 156
Chapter 5 Operation Precautions
5.1 Precautions before use
5.1.1 For safe and effective use, please read the user manual carefully before operation.
5.1.2 Check to ensure that the device is in good condition.
5.1.3 The device shall be placed on a flat surface, and moves gently to avoid strong vibration or
shock.
5.1.4 Check to ensure that the lead cables are correctly connected, and the device grounding is
correct.
5.1.5 The AC frequency and voltage should comply with the requirements, and enough current
capacity should be guaranteed.
5.1.6 When using the battery for power supply, check to ensure that the battery voltage and
battery status are in good condition, and the battery has enough power.
5.1.7 When the device is used together with other equipment, all devices and equipment should
be equipotential grounded in order to protect the user and operator.
5.1.8 Install the device where easily grounded in the room. Do not allow the patient and
patient-connected lead cables and electrodes to come into contact with other conductor parts,
including the earth or a hospital bed.
5.1.9 Clean the lead cable with neutral solvent. Do not use alcohol-based cleaners or gemicides.
5.1.10 Ensure that the device is running within the normal ambient temperature range of 5℃ to
40℃. If the device is stored at a higher or lower temperature, leave it in the operating
environment for approximately 10 minutes before use in order to ensure the normal work.
5.2 Precautions during operating
5.2.1 The printing can be started after the ECG waveform is stable.
5.2.2 During using, the doctor should observe the patient carefully and cannot leave the
operating site. If necessary, turn off the power or remove the electrode to ensure patient safety.
5.2.3 The patient and the device can only be connected via lead cables through the electrodes, in
order to avoid patient touches other parts of device or conductors.
5.2.4 Patient can not move during operating.
5.2.5 Maintenance or repair to the device or accessory is not allowed during using.
5.3 Precautions after use
5.3.1 Set the states of all functions to initial states.
5.3.2 Cut off the power, gently remove the electrodes and limb clips, then remove the lead cables,
do not pull with force.
5.3.3 Clean the device and all accessories, and store them for the next use.
14

Publicité

loading

Ce manuel est également adapté pour:

33232