exceeds 60°C. Do not heat the battery or throw it into fire, water and avoid splashed
by water.
Do not puncture, hammer or strike the battery or destroy it by other ways, otherwise
it will cause battery overheat, smoke, deform or burn dangers.
Keep away from the battery when it appears leakage or emitting unpleasant smell. If
the battery electrolyte leaks onto the skin or clothes, clean with water immediately. If
the electrolyte accidentally enters your eyes, do not rub your eyes, immediately clean
with water and see a doctor.
If the battery reaches its service life, or battery smell, deform, discolor or distorted
appears, please stop using the battery and dispose it in accordance with local
regulations.
9.2 Recording paper
In order to ensure the quality of the ECG waveform, please use the high-speed thermal
recording paper supplied or specified by the company. If you use unspecified recording paper,
the recorded ECG waveform may be blurred, faded, and the paper feeding may not be smooth.
This may even increase the wear of the device and shorten the service life of important parts
such as the thermal print head. For information on how to purchase such recording paper, please
contact your dealer or the company. Please be careful!
9.2.1 When using recording paper, it is absolutely not allowed to use recording paper with wax
on the surface or in grayish/black color. Otherwise, the wax will stick to the heating part of the
print head, resulting in abnormal work or damage of the print head.
9.2.2 High temperature, humidity and sunlight may cause the recording paper to change color.
Please keep the recording paper in a dry and cool place.
9.2.3 Please do not place the recording paper under fluorescent light for a long time, otherwise it
will affect the recording effect.
9.2.4 Please do not to put the recording paper together with the PVC plastic, otherwise the color
of recording paper will change.
9.2.5 Please use the recording paper with specified dimension. Recording paper that does not
meet the requirements may damage the thermal print head or silicone rubber roller.
9.3 Maintenance after use
9.3.1 Turn off the device.
9.3.2 Unplug the power cord and lead cables. Hold the header of plug to disconnect, and do not
pull the cable with force directly.
9.3.3Clean the device and accessories, cover them up to against dust.
9.3.4 Store the device in a cool and dry place, avoid strong vibration when moving.
9.3.5 When cleaning the device, do not immerse it in the cleaner. Power supply must be cut off
before cleaning. Use neutral detergents for cleaning. Do not use any detergent or disinfectant
containing alcohol.
9.4 Lead cables and electrodes
9.4.1 The connectivity of the lead cable can be detected by the multimeter. Check whether each
wire of the lead cable is in good contact according to the following table. The resistance of each
wire from the electrode plug to the corresponding pin in the lead cable plug should be less than
37