Old inferolateral MI
45
ST depression, mild anteroseptal myocardial ischemia
46
ST depression, mild anterior myocardial ischemia
47
ST depression, mild extensive anterior myocardial ischemia
48
ST depression, mild apical myocardial ischemia
49
ST depression, mild anterolateral myocardial ischemia
50
ST depression, mild high lateral myocardial ischemia
51
ST depression, mild inferior myocardial ischemia
52
ST depression, mild inferolateral myocardial ischemia
53
ST depression, anteroseptal myocardial ischemia
54
ST depression, anterior myocardial ischemia
55
ST depression, extensive anterior myocardial ischemia
56
ST depression, apical myocardial ischemia
57
ST depression, anterolateral myocardial ischemia
58
ST depression, high lateral myocardial ischemia
59
ST depression, inferior myocardial ischemia
60
ST depression, inferolateral myocardial ischemia
2.3 Intended use
The intended use of the Automated Measurement&Interpretation function is shown as
below:
Application
and diagnosis
Population
Application
site
Accuracy
Others
3. Algorithm description
This section describes the algorithm, formulas and judgment conditions for interpretation
items related to functions of ECG automated measurement and automated interpretation.
The 12-lead sync ECG waveform passes through the filter (AC, EMG, DFT (if has, and
open)) into the module of automated measurement and automated interpretation.
The module of automated measurement and automated interpretation mainly includes
process of find the cardiac impulse location, find the beginning/end for each wave, amplitude
calculation, parameters calculation, and interpretations judgment based on known parameters.
To detect the abnormal of heart of human body, examination items refer to
above description
Teenagers and adults, age range: 12-87
hospitals
The accuracy of this function is reflected by the balance performance of
sensitivity and specificity.
This function does not generate any alarm when using, so it should be
operated by professional or trained personal.