Problem
Rapid alarm condition, and the
(yellow) BATT light illuminates
intermittently and FreeStyle shuts
down.
Intermittent alarm condition, and the
(red) alarm light illuminates
intermittently.
Intermittent alarm condition, and the
(red) alarm light illuminates
continuously.
Unit does not start on battery power.
Delay in recharging internal battery.
Unit alarms while in automobile and
connected to the DC outlet.
All other problems.
MN131-1 rev C
Probable Cause
Battery voltage is too low to operate
the FreeStyle unit.
Breathing rate has exceeded the
capacity of the FreeStyle unit.
A general malfunction has occurred.
Unit may be hot or cold if left
outdoors such as in an automobile.
Internal battery exceeds charging
temperature.
No power to the unit if battery
depleted and DC outlet not charging.
Solution
Connect to DC or an AC outlet
immediately.
Reduce activity, and then turn unit off and
back on again to reset unit. If necessary,
change to another source of oxygen as
available and contact your Equipment
Provider.
Change to another source of oxygen as
available, and contact your Equipment
Provider.
Allow the unit to reach normal operating
temperature, which may take several
minutes if exposed to temperature
extremes. Temporarily connect your
Universal AC/DC power supply to the
unit's power inlet and power source, as
needed, to reset the unit's internal
battery.
Unit may be operated; however, charging
may not resume until battery temperature
is reduced. (See page 9 notes.)
Disconnect the Universal AC/DC power
supply from the automobile outlet, restart
the automobile, and then reconnect the
power supply into the automobile DC
outlet to reset the breaker within DC
power supply.
Change to another source of oxygen as
available, and contact your Equipment
Provider.
EN-35