DMC plus and Digital Twin System Hands, hand size 7
1. Shorten the cable to approx. 5 cm after it has exited the bore hole of the
tube chassis of the System Hand (fig. 26). Separate the ends of the indi-
vidual strands of the cable, remove 2 mm of the insulation and tin (fig. 27).
2. Use the 711M12 Soldering Jig to solder on the plug. The marking in
Fig. 31 shows the position of the red strand of the hand cable.
3. Press in the control electronics, tighten the screws (fig. 17-15) and attach
the plug. Perform a function check. If necessary, reposition the 10V40
MyoWrist 2Act once again in relation to the 8E41 Ottobock System
Electric Hand (fig. 32).
5 Legal information
All legal conditions are subject to the respective national laws of the country
of use and may vary accordingly.
5.1 Liability
The manufacturer will only assume liability if the product is used in accor-
dance with the descriptions and instructions provided in this document.
The manufacturer will not assume liability for damage caused by disregard
of this document, particularly due to improper use or unauthorised modi-
fication of the product.
5.2 CE conformity
This product meets the requirements of the European Directive 93 / 42 / EEC
for medical devices. This product has been classified as a class I device
according to the classification criteria outlined in Annex IX of the directive.
The declaration of conformity was therefore created by the manufacturer
with sole responsibility according to Annex VII of the directive.
Ottobock | 25