Chapter 4: Alarms and Troubleshooting

Alarm and Indicators
The device has an audible alarm and three LED indicators, as shown below.
Note: All EverFlo alarms are low priority alarms. The alarm system should be verified before use and
between users by service personnel in accordance with the EverFlo service manual.
Audible Alarm / Colored LED
All 3 LEDs illuminate continuously
and the Audible Alarm is
sounding continuously.
The Audible Alarm is sounding
continuously. None of the LEDs
are illuminated.
Red LED illuminates continuously
and the Audible Alarm is
sounding continuously.
NORMAL
Green Power On Indicator
Yellow Check System Indicator
Red Alarm Indicator
Possible Cause
The device has
detected a system
malfunction.
The device is turned on
but is not operating.
Often this indicates
that the device is not
plugged in or there is a
power failure.
The device has
detected a system
malfunction.
EverFlo / EverFlo Q User Manual
Your Action
Immediately turn off the device,
connect to a back up oxygen
source, and call your home care
provider.
Check the power outlet and verify
that the device is plugged in. If the
problem continues, connect to a
back up oxygen source and call
your home care provider.
Immediately turn off the device and
wait 5 minutes. Restart the device. If
the condition persists turn the unit
off, connect to a back up oxygen
source, and call your home care
provider.