Technical Alarms - Philips RESPIRONICS SimplyGo Mini Mode D'emploi

Masquer les pouces Voir aussi pour RESPIRONICS SimplyGo Mini:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Technical Alarms

Visual, Audio
Indicators
1 beep
that repeats every
16 seconds
AMSR
Description
No Breath Alarm
This alarm occurs when
a breath is not detected
for a period of 2 minutes
or more� Press the yellow
Alarm silence button on
the screen to temporarily
silence the alarm, and
the 2nd screen shown at
left appears� Or, press the
Home button on the device
to return to the Home
screen, and the 3rd screen
will appear with the No
Breath (NB) alarm symbol
in the upper left corner to
indicate the active alarm
state�
This alarm resets itself
as soon as a breath is
detected� If the device is
on battery power and no
breath is detected after
approximately 15 minutes,
the device shuts down
to conserve power� (This
alarm is inactive for ten
minutes after startup�)
Note: Upon startup,
this alarm could take a
maximum of 12 minutes to
activate from the time the
alarm condition is present�
Otherwise, it can take up to
2 minutes to activate�
oxygenconcentratorstore.com
SimplyGo Mini User Manual
What To Do
Check the
connection from
the cannula to the
device�
Make sure the nasal
cannula is properly
positioned on your
face and that you are
breathing through
your nose�
Make sure the
cannula tubing is not
kinked or obstructed�
1-877-774-9271
25

Publicité

Table des Matières
loading

Table des Matières