Visual, Audio
Indicators
2�0
AMSR
Description
Alarm Symbol
This gray symbol appears
when you have pressed
the alarm silence button�
Pressing this button will
enable the alarm audible�
Attention
This yellow symbol
appears on the status bar
when corrective action is
required�
Flow Control Setting
The large number on
the screen shows the
oxygen flow setting (in
this example, it is 2)� This
number can range from 1
to 5 in increments of 1�
oxygenconcentratorstore.com
SimplyGo Mini User Manual
What To Do
Refer to the
associated alarm
indicator also being
displayed�
If the device is
operating in the
carrying case, make
sure it has been
placed correctly
in the case so that
the outlet vents
are not blocked�
See the Carrying
Case section of this
manual�
Allow the device
to warm up� If the
alarm continues,
contact your
equipment
provider�
If this is your flow
setting, no action is
needed�
If this is not your
flow setting,
press the + button
to increase this
number or press the
- button to decrease
the number�