Indicators
1 beep
that repeats every
16 seconds
AMSR
32
Description
Depleted Battery Alarm
This alarm occurs when
approximately two
minutes of battery life
remains�
Press the yellow Alarm
silence button on the
screen to temporarily
silence the alarm, or press
the Home button on the
device to return to the
Home screen, and the
2nd screen will appear
with the Depleted Battery
symbol flashing on screen�
Note: This alarm could
take up to 46 seconds to
activate from the time
the alarm condition is
present�
Note: The device will
automatically power
down if the battery
remains depleted for
more than 90 seconds
and no external power
source is connected�
Alarm Silence Symbol
This yellow symbol
appears when an event
occurs that generates an
audible alarm� Pressing
this button will silence
the audible alarm until a
different alarm occurs or
until you disable the alarm
silence (see below)�
oxygenconcentratorstore.com
What To Do
Replace the battery
or connect to a
power source�
Refer to the
associated alarm
indicator also being
displayed�
1-877-774-9271