Powering On the Device

1�
You can press the Power button once or simply
touch the black screen on the device to quickly
view your settings� The screen shown here will
appear with the following information displayed:
O
setting
2
Battery capacity
Number of hours the device has been used
Software version
The screen will automatically shut off again after 5 seconds�
2�
If you want to fully power on the device, press
the Power button a second time, within 5 seconds of the first press�
This second press helps prevent unwanted starts or inadvertent
changes in oxygen flow� The Philips Respironics splash screen
appears briefly followed by the Home screen, both shown below�
Every time it starts up, the device starts producing oxygen to meet
the specified setting� It may take up to 20 minutes to reach the
specified purity levels for the setting� It is safe to begin breathing
from the SimplyGo Mini device immediately� Breathing from the
device will cause it to reach oxygen purity levels faster�
Note: When it is turned on, the device will automatically start at the oxygen
flow rate setting used when the device was last turned off� As a
precaution, each time you start the device, verify the flow setting�